Stability

In order to evaluate the stability of the spectrophotometer, 48 experiments were done without delay.

This experiment was repeated twice on D1.

The variation for RGB is less than 0.1%.

D1_1.svg

D1_2.svg

The experiment was redone using a delay of 600s between each acquisition. The main variation is probably due to temperature changes. We observe that the changes can now reach up to 0.5% and are depend on the color.

D1_3.svg

In order to confirm the importance of the temperature, the experiment was repeated with the acquisition of the temperature.

D1_4.svg

D1_4temp.svg

We can clearly see an inverted correlation between the red variation and the temperature.